|
|
What is the mechanism of the Sophysa Adjustable Pressure Valves? |
|
|
|
The
principle of the Sophy® and Polaris® adjustable
pressure valves is based on the pressure variation exerted
on a ball by a flat semi-circular spring at various points
of its curvature. At each contact point of the spring
with the ball, corresponds a resistance or differential
pressure.
The ball-in-cone mechanism situated at the extremity
of the inlet connector, ensures a perfect anti-reflux
function and is insensitive to temperature variations.
Tests have demonstrated that Sophy® and Polaris® valves
did not exhibit any reflux for outlet-inlet differential
pressure of up to 200 mmHg, i.e. more than 2700 mmH2O.
The spring is connected
to a magnetic rotor whose position can be modified
non-invasively by using
an adjustment
magnet.
The rotor of Polaris® features a patented self-locking
mechanism which prevents unintentional dysadjustment
during everyday life and MRI exposure.
|
|
|
Out of which
material are the Sophysa Adjustable Pressure Valves
made? |
|
|
|
| The
chamber of the valve : |
Polysulphone |
| The
ball : |
Synthetic
Ruby |
| The
valve connectors : |
Medical
grade 316 L stainless steel |
| The
2 micro-magnets of the valve rotor : |
Samarium
Cobalt |
Sophysa Adjustable Pressure Valves are latex free. Catheters
included in Sophysa shunt kits are made of silicone.
|
|
|
What is
the precision of the calibration of the Sophysa Adjustable
Pressure Valves? |
|
|
|
All
along the pressure range of the SM8 and SPV models,
the precision of calibration is within -10/+15 mmH2O
from the nominal value. The flat semi-circular spring
ensures a precise and reproducible functioning pressure
for each position of the rotor. It is resistant to
ageing and maintains constant performance over time.
|
|
|
What are
the operating pressures of the Sophy® and Polaris® valves? |
|
|
|
Sophy® adjustable
valves
- 50, 110, 170 mm H2O for the model SM3
- 30, 50, 70, 90, 110, 140, 170, 200 mm H2O for the model
SM8
- 50, 75, 100, 125, 150, 180, 220, 300 mm H2O for the
model SM8-300
- 80, 120, 150, 190, 230, 270, 330 400 mm H2O for the
model SM8-400
- 10, 25, 40, 60, 80, 100, 120, 140 mm H2O for the model
SM8-140
Polaris® adjustable valves
- 30, 70, 110, 150, 200 mm H2O for the model SPV
- 50, 100, 150, 220, 300 mm H2O for the model SPV-300
- 80, 150, 230, 330, 400 mm H2O for the model SPV-400
- 10, 40, 80, 110, 140 mm H2O for the model SPV-140
|
|
|
Why are
the Sophy® and Polaris® valves stiff? |
|
|
|
The
Sophy® and Polaris® valves are stiff, because
the mechanism has to be protected against shock, pumping
operations or needle punctures. Nevertheless its anatomic
shape is especially adapted to the skull's convexity.
This non-distorting structure is insensitive to subcutaneous
pressure variation, unlike silicone valves.
|
|
|
Can we use
a anti-siphon device with the Sophy® and Polaris® valves? |
|
|
|
Yes.
It will have to be placed distally to the valve and
the valve should be set towards the lowest pressures.
|
|
|
Why don't
you supply a Sophy® or Polaris® valve with
an On-Off system? |
|
|
|
The
On-Off systems available on the market do not avoid
the risk of an accidental closure.
|
|
|
What about adjustment
of the Sophy® or Polaris® valve before implantation? |
|
|
|
It
is advisable to adjust the valve before implantation.
Before opening the sterile packaging, place the rotor
in the position selected for implantation, by placing
the adjustment magnet over the valve in the position
of the packaging shell designed for this purpose, with
its north pole placed to the left of the direction of
the flow. Turn the magnet in order to adjust the rotor
to the desired position.
This pre-implantation adjustment avoids contact with
the scar of the implantation site, reduces the risk
of infection and ensures better patient comfort.
|
|
|
Do you recommend a pressure
test before implanting a Sophy® or Polaris® valve? |
|
|
|
No.
As each valve has been individually calibrated and
controlled with specific high precision equipment,
it is unnecessary to perform any additional tests prior
to implantation.
Per-operative testing increases operating time and
infection risk and thus should be avoided.
Neurosurgeons that use regularly the Sophy® and Polaris® adjustable
pressure valve have now evidence of the reliability of
the valve.
|
|
|
Are there any precautions
to be taken during implantation procedure of Sophy® or
Polaris® valve? |
|
|
|
- Sophy® or Polaris® have to be thoroughly
flushed from air bubble, either with the patient
CSF or with
sterile water.
- The depth of implantation MUST NOT exceed 10 mm
below the skin surface for Sophy® and 8 mm for Polaris®.
- Before closing the implantation site, the arrow
moulded on the upper surface of the Sophy® and
Polaris® MUST
be visible and MUST point the flow direction of the
CSF.
|
|
|
Is there any recommended
pressure setting for Sophy® or Polaris® at
the time of implantation? |
|
|
|
More
and more neurosurgeons advocate initial pressure setting
toward highest pressure whatever the etiology in order
to reduce the risk of overdrainage related complications
:
Miyake et al., in Neurosurgery, Vol. 40, No. 5, May
1997
« Our strategy for the initial
setting of the programmable valve may have also
contributed
to the favorable results. We set the initial
pressure at the highest
level below the preoperative ICP, which may have decreased the
incidence of overdrainage-related complications, even in patients
with with high pressure hydrocephalus. »
Lee et al., in Child’s Nerv Syst (2002), 18:533-563
« The use of programmable shunt devices
has been strongly recommanded for all types of
hydrocephalus
and especially
for NPH, but vigorous efforts should be invested
in finding the optimal pressure for each patient
to get
the best results from shunting. Our suggestion
for adjustment is to set the initial valve
pressure at
a level 10-30 mm H2O lower than their CSF pressure
and decrease the pressure by 30 mm H2O every
3 weeks (could be shortened to 1 or 2 weeks, depending
on symptoms
or CT findings) until the symptoms have
improved and the ventricles have normalized. We have found
that
this method is effective and safe in reducing
the risk of complications related to overdrainage
syndrome. »
|
|
|
Is it possible to implant
the Sophy® valve as a lumbo-peritoneal shunt? |
|
|
|
Yes,
it is. The Sophy® valve placed on the flank region
should be fixed over a bony surface like lower ribs
in order to facilitate its localiation for adjustment.
The catalog number for a complete lumbo-peritoneal
Sophy® shunt kit preattached is SM8-2040.
Polaris® valve is contra-indicated for lumbo-peritoneal
shunting, since the depth of implantation is likely to
exceed 8 mm from the skin surface.
|
|
|
Is there any difference
in flow regulation between skull and chest implantation
of a Sophy® valve? |
|
|
|
The
Sophy® valves can be implanted either on the skull
or in the infra clavicular fossa without any difference
in the flow regulation and drainage performances.
In case of chest implantation, the valve should be
sutured, by using the integrated fixation holes, to
the underlying
fascia in order to prevent valve rotation on its axis.
Polaris® valve is contra-indicated for
chest implantation
since the depth of implantation is likely to
exceed 8 mm below the skin surface.
|
|
|
How should one read
the pressure setting of the Sophy® valve on X-Ray
film? |
|
|
|
The
main pressure settings of the Sophy® adjustable
pressure valve can be easily determined both visually
and radiographically by means of a standard identification
code. From the inlet connector, on the right side of
the valve is a sequence of one, two and three dots
corresponding to the 3 main pressure settings.
When the right micro-magnet of the rotor is opposite
:
- One dot = Low pressure setting
- Two dots  = Medium pressure setting
- Three dots   =
High pressure setting
The intermediary pressure settings can be assessed
by the position of the right micro-magnet in relation
to the
dots.
|
|
|
How should one read
the pressure setting of the Polaris® valve on X-Ray
film? |
|
|
|
Each
pressure setting of the Polaris® adjustable pressure
valve can be easily determined both visually and radiographically.
From the inlet connector, on the right side of the
valve is a sequence of 5 dots, each dot corresponds
to one of the 5 pressure settings.
When the right micro-magnet of the rotor is opposite
:
- To nearest dot from the inlet connector = Lowest
pressure setting (N°1)
- To the furthest away dot from the inlet
connector = Highest pressure setting (N°5)
|
|
|
What is the method of
ensuring from an X-ray that an implanted Sophy® or
Polaris® valve
is correctly in place (the relief arrow on
the skin side), or upside down (relief arrow on the
skull side), regardless of the side from which the
X-ray film is read? |
|
|
|
A
valve implanted on the right side of the skull is correctly
placed if the radio-opaque dots indicating the
pressure settings are pointing towards the patient's
nose.
A valve implanted on the
right side of the skull is incorrectly placed (upside
down) if the radio-opaque
dots indicating the pressure
settings are pointing towards the patient's neck.
|
|
|
What about adjustment
if the Sophy® or Polaris® valve has been implanted
upside down (moulded arrow seated on skull side instead
of skin side)? |
|
|
|
- The direct reading with the compass will be inverted
: a pressure setting in LP will be read HP with
the compass; a pressure setting HP will be read LP.
- For adjustment, the north pole of the magnet
will have to be placed to the right of the direction
of
the flow.
|
|
|
What could be the main
reasons for difficulty or failure of the pressure adjustment
on implanted Sophy® or Polaris® valve by a
physician? |
|
|
|
By
experience the main reasons can be listed as follows:
- Unsuitability of the
adjustment kit with the implanted valves.
Be sure that the Sophy® adjustment kit (Cat. Nb
= RS5) is used for a Sophy® valve, and a Polaris® adjustment
kit (Cat. Nb = PAK) is used for a Polaris® valve.
- Incorrect manipulation of the adjustment kit
( see IFU).
- The valve has been implanted upside down.
Check on the X-ray the correct implantation of the
valve (see questions N°10, N°11 and N° 17,
N°18).
- Valve implantation is too deep beneath the skin
surface.
At the time of implantation, the depth below the
skin should not exceed 10 mm for Sophy® and 8
mm for Polaris® valve.
- The rotor of the valve is blocked by biological
deposit.
|
|
|
Is it possible to test
the patency of the shunt once the Sophy® or Polaris® valve
is implanted? |
|
|
|
Yes.
A ventricular catheter with a reservoir or a valve
with integrated reservoir is necessary to check the
patency
of the shunt.
|
|
|
Can the patient feel
some noise from the Sophy® or Polaris® valve
implanted on the skull? |
|
|
|
Yes.
This is quite frequent and whatever the disturbance
it may cause, it also means that the shunt is functioning
well.
The reason for this is the rotation of the ruby ball
on its seat and the turbulence induced by CSF flow
through
the rigid structure of the valve. These noises are
amplified and transmitted by the skull to the cochlea.
These noises mainly occur when the patient is changing
position leading to a temporary increase of the CSF
flow through the shunt.
|
|
|
What is the incidence
of MRI on the pressure setting of the Sophy® and
Polaris® Valve? |
|
|
|
- The pressure setting of the Sophy® adjustable
valve may be altered when exposed to strong magnetic
fields.
It is important to mention that this is true for
other types of adjustable pressure valves on
the market,
EXCEPT Polaris®.
Outside the 0.5 mT safety line, which encircles the
MRI scanner, there is no disadjustment.
The Polaris® adjustable pressure valve is the ONLY
adjustable pressure valve with no disadjustment while
exposed to a strong magnetic field such as 3 Tesla MRI.
- The micro-magnets included in every adjustable pressure
valves are responsible for the artefacts on the images.
The artefacts induced by Sophy® and Polaris® valves
are similar. If critical areas during the follow-up
of the patient are expected to be obscured by artefacts,
knowledge of the expected degree of signal void and
image distortion may influence surgical decision regarding
site of implantation.
- There is no damage of the Sophy® and Polaris® valve
mechanism even after multiple MRI exposure, meaning
that the Sophy® and Polaris® valves remain
adjustable. This has been demonstrated with exposure
to 3 Teslas.
|
|
|
Is it possible that
the Sophy® and Polaris® valve be dislocated
from its site under MRI? |
|
|
|
NO !!!
cf : study of Ortler, Neurosurgery, Vol.40, No 5, May
1997 :
"… the force acting on the
valves during MRI is not strong enough to dislocate
or
rotate a valve."
"…
the torque acting on a Sophy® Valve corresponds
approximately to the torque acting on a Mc Fadden
aneurysm clip."
|
|
|
Is there any heating
effect from the Sophy® or Polaris® valve under
MRI? |
|
|
|
No
heating effect at the valve fixation site has been
experienced by volunteers.
|
|
|
What are the relationships
between operating pressure of the valve and intracranial
pressure? |
|
|
|
The
operating pressure does not correspond to intracranial
pressure, but is only the resistance of the valve
to CSF flow. Catheters also add resistance to the CSF
flow.
Drainage occurs as a result of positive pressures
(intraventricular pressure or ICP, weight of the height
of the CSF column)
and negative pressures (cerebral resistance to flow
/ poor compliance, valve operating pressure, catheters
length and diameter, intra-abdominal pressure*). Thus
there is no evidence that the ICP should have the same
value as the operating pressure of the valve.
*in case of a ventriculo-peritoneal shunt
|
|
|
What to recommend for
a patient with a Sophy® SM8 or Polaris® SPV
valve set to 110 mm H2O who complains of headache when
he is standing, which resolves when he lies down, and
in whom a CT scan shows normal size ventricles? |
|
|
|
The
symptoms described are probably related to overdrainage.
The absence of slit ventricles on a CT scan could
suggest that this is not the case, but in the presence
of contradicting
signs/symptoms, decisions should always be based on
the clinical signs. The development of clinical signs
often
precedes the effects of overdrainage on ventricular
volume and therefore the appearances seen on a CT scan.
In the case described, we would therefore increase
the valve operating pressure (moving from medium
pressure = 110 to high pressure = 200) in order to
reduce drainage
and we would monitor changes in the patient's symptoms
and, if necessary set later the valve to an intermediary
pressure.
|
|
|
What to recommend for
a patient with a Sophy® SM8 or Polaris® valve
set to 70 mm H2O in whom an early sub-dural haematoma
has been diagnosed? |
|
|
|
Development
of a sub-dural haematoma in a patient with hydrocephalus
carrying a shunt probably indicates a well-known side
effect, which is a consequence of overdrainage.
A CT scan would probably show slit ventricles.
In the case described the valve is set to 70 mm H2O.
This is a relatively low operating pressure and we
would therefore recommend to increase the valve
pressure. The
neurosurgeon could also decide immediate drainage
of the heamatoma , depending on the patient's condition.
|
|
|
What is the slit ventricle
syndrome due to, in a patient carrying a shunt with
a valve? How would this be remedied in a case in which
the patient was carrying a Sophy® SM8 or Polaris® SPV
set to 200 mm H2O? |
|
|
|
The
slit ventricle syndrome is due to overdrainage, the
major risk of which is development of a sub-dural haematoma.
We would therefore increase the valve pressure.
In the case described, the valve is set to 200 mm
H2O, the maximum pressure for an SM8 or SPV valve.
It is therefore
impossible to increase the pressure of this valve
and it will therefore have to be replaced by a specific
SM8 or SPV valve with an operating pressure of
up to 300
(SM8-300, SPV-300) or even 400 mm H2O (SM8-400, SPV-400).
|
|
|
What precautions does
a patient carrying a Sophysa Adjustable Pressure
Valve have to take concerning everyday life? |
|
|
|
Magnetic
fields generated by airport security doors, microwave
oven, mobile phone, high tension wire, TV, etc...,
can potentially alter the pressure setting of the Sophy® valve.
The permanent magnets, domestic, in toys, in audio
headphones, in loud speakers, and electromagnetic
field created by
electric motor from shaver, hair dryer, clipper, while
turned on, etc…can also modify the Sophy® pressure
setting. Thus in order to prevent any risk of pressure
alteration, they must not be brought close to the valve
site. Direct shocks on the implanted site of the valve
can also be responsible of disadjustment.
It is important to mention that this is true for the
other types of adjustable pressure valves on the market,
EXCEPT Polaris®.
The patented Self-locking
System of the Polaris® adjustable
pressure valve makes it insensitive to any
everyday magnetic fields unintentionaly approached
and also to direct shocks.
|
|
|
Is it possible to place
the Sophy® or Polaris® valve between a ventricular
catheter and an external drainage system? |
|
|
|
Using
an external drainage system with a valve should be
avoided. In case of temporary externalisation of a
ventriculo-atrial / peritoneal shunt with a valve,
the follow-up is identical to a conventional external
drainage. It will have to be performed by people having
the necessary skills and under the responsibility of
a physician.
|
|
|
Are there any contra-indications
in performing laparoscopic surgery on patients with
Sophy® or Polaris® valve ? |
|
|
|
There
are no contra-indications for laparoscopy on patients
with Sophy® or Polaris® valves: Inflating
CO2 pressure ranges from 10 to 15mmHg, which is
far below the anti-reflux performance of valves tested
up to 200 mmHg.
However, the shorter the laparoscopic surgery the
better, since the abdominal pressure will reduce
the
ventriculo-peritoneal drainage during inflating.
It can also be recommended
to decrease the pressure setting of the adjustable
valve during surgery.
|
|
|
How is it possible to
trace the Sophy® and Polaris® valves ? |
|
|
|
All
the Sophy® and Polaris® valves are manufactured
with an individual serial number etched on the inferior
face of the valve casing.
Moreover, each Sophy® and Polaris® valve
is packaged with a Patient Identification Card (PIC)
and a set of
stick-on labels mentioning the Cat. Number, Lot Number
and Serial Number.
The PIC should be carried by the patient at all times,
providing the information of the implanted device (reference,
pressure setting, …).
|
|
|
Can a Sophysa valve
be re-sterilised ? |
|
|
|
No.
Our products are sterilised with ethylene oxide. They
are for single use only and must not be re-sterilised
or used if the packaging is opened or damaged, or beyond
the expiry date.
If the packaging has been opened and the valve unused,
it has to be returned to SOPHYSA for re-sterilisation.
|
Privacy Policy, Legal
notice, Contact
us
|